Research Article
Stability-Indicating HPLC Determination of Trandolapril in Bulk Drug and Pharmaceutical Dosage Forms
Table 3
Precision study of the proposed method.
|
Concentration (g/mL) | Intraday precision | Interday precision | Conc. found Mean ± SD | % CV | Conc. found Mean ± SD | % CV |
| 4 | 3.937 ± 0.068 | 1.74 | 4.008 ± 0.039 | 0.98 | 12 | 11.949 ± 0.088 | 0.74 | 12.011 ± 0.108 | 0.90 | 22 | 21.853 ± 0.102 | 0.47 | 21.731 ± 0.151 | 0.70 |
|
|