Research Article

Development and Validation of an HPLC Method for Simultaneous Assay of MCI and MI in Shampoos Containing Plant Extracts

Table 6

Results of robustness.

VariationSpecific conditionMIMCI
RSD (%) for peak areaRSD (%) for content in shampooRSD (%) for peak areaRSD (%) for content in shampoo

Flow rate (mL/min)0.90.40.40.40.3
1.0 (normal)0.60.40.40.4
1.10.60.50.60.3

Percentage of methanol in mobile phase (%)290.80.60.40.6
30 (normal)0.60.50.40.3
310.50.40.50.3