Extraction recovery of cortisol and internal standard (IS) from saliva.

Nominal concentration
(ng/ml)
Spiked-before-extractionSpiked-after-extraction†Recovery
()
CortisolMeanSDMeanSD

0.59737610291595
1.51652131191017886
5049728481453252435893
901097629774115761408795
IS 20103554159231096301799094