Intra-and inter-run precision and accuracy of cortisol assay.

Nominal level
(ng/ml)
Intra-day (n=10)Inter-day (n=20)
Mean (SD)
measured
level
CV
()
Bias
()
Mean (SD)
measured
level
CV
()
Bias
()

0.50.52 (0.05)9.03.20.51 (0.04)8.42.2
1.51.42 (0.09)6.0-5.51.47 (0.11)7.7-2.0
5052.77 (1.49)2.85.153.27 (2.61)4.96.5
90100.79 (2.38)2.412.099.23 (3.86)3.910.3