Research Article

A Validated Liquid Chromatographic Method for Berberine Analysis in Tissue and Application

Table 2

Precision and accuracy of the method (n = 6).

TissuesConcentration (μg/g)Within-day precision RSD (%)Between-day precision RSD (%)Recovery (%)

Liver0.255.305.8792.66
1.252.372.53100.1
10.03.934.00102.2

Kidney0.100.981.06107.0
0.251.742.4892.82

Spleen0.102.823.63106.7
0.251.742.32103.0