Research Article

Long-Term Stability of Hydromorphone in Human Plasma Frozen at −20°C for Three Years Quantified by LC-MS/MS

Table 2

Intra-assay accuracy and precision of the applied analytical method.

QC sampleNominal concentration (ng/mL)Mean ± SD (ng/mL)%RSD%REn

High concentration20.019.8 ± 0.472.35−0.905
Middle concentration2.502.54 ± 0.0552.161.605
Low concentration0.310.32 ± 0.04514.02.405

SD: standard deviation; %RSD: relative standard deviation; %RE: relative error from nominal concentration.