Analytical Validation of Smartphone Spectroscopic Technic Used in an Educational Kinetic Study

Table 4

Recapitulation of validation steps.

CharacteristicTestDescription of the test

Random characterShapiro–Wilk test conformFor the risk α = 5%, R, and 12 repetitions, the number of sequences R = 8 is between higher and lower values.
SpecificitySpecific methodResponse of solutions prepared in water and in reaction mixture are equal.
PrecisionCochran test conformThere is no aberrant responses in the 12 measurements; the fraction of higher variance to the sum of variances is less than Cochran critical vale at the risk α = 5% and α = 5%.
LinearityFisher test conformFisher test showed that the fraction of calculated residuals and experimental ones is inferior to the tabulated Fisher value for 5 levels repeated 3 times.
Calibration functionA = 1000 C + 0.002Least square regression is involved to determinate the slop and the intercept of the calibration curve.
Correlation coefficient0.9999This value represents the fraction of the variation in one variable that may be explained by the other variable.
Limit of detection mol·L−1Statistical test of comparison of the response at the value 0 becomes significant. The limit of detection is determined with a risk of 0.05%.
Limit of quantification mol·L−1Statistical test of comparison of the response at the value 0 becomes significant. The limit of quantification is determined with a risk of 0.05%.
Uncertainty mol·L−1Because of the difference between the real value and the measured one, a degree of uncertainty will pertain to measurement. Uncertainty is the absolute range in which measured value can be accepted.