Research Article
Method Validation and Measurement Uncertainty (MU) Evaluation on Enrofloxacin and Ciprofloxacin in the Aquatic Products
Table 2
Data for calibration curve of enrofloxacin and ciprofloxacin.
| The analyte | Linear concentration (μg/mL) | Peak area | Equation of calibration curve | Correlation coefficient (R2) |
| Enrofloxacin | 0.000 | 0.00 | y = 81320x − 84.917 | 0.9998 | 0.002 | 140.28 | 0.010 | 676.10 | 0.050 | 3959.51 | 0.100 | 7884.95 | 0.200 | 16267.59 |
| Ciprofloxacin | 0.000 | 0.00 | y = 37049x − 42.688 | 0.9997 | 0.002 | 61.04 | 0.010 | 304.15 | 0.050 | 1805.37 | 0.100 | 3569.70 | 0.200 | 7415.34 |
|
|