Research Article

The Determination of Polymyxin B in Critically Ill Patients by the HPLC-MS/MS Method

Table 3

Intrabatch and interbatch precision and accuracy for PB in plasma.

ComponentNominal conc. (ng/mL)Intraday (n = 5)Interbatch (n = 3)
Found conc. (ng/mL) mean (SD)Acc. (%)CV (%)Found conc. (ng/mL) mean (SD)Acc. (%)CV (%)

PB1/PB1-I
LLOQ100100 (0.005)99.95.494 (0.005)94.05.0
QCL200188 (0.007)94.23.5190 (0.012)94.86.3
QCM11000934 (0.039)93.44.1953 (0.059)95.36.2
QCM240003972 (0.082)99.32.14128 (0.274)103.26.6
QCH70007036 (0.040)100.50.67105 (0.328)101.54.6

PB2/PB3
LLOQ11.912.7 (0.001)106.86.612.7 (0.001)107.06.4
QCL23.724.0 (0.002)101.36.624.7 (0.002)104.37.2
QCM1118.6113.8 (0.002)96.02.1114.6 (0.005)96.64.1
QCM2474.3452.4 (0.016)95.43.4466.0 (0.022)98.34.7
QCH830.0801.0 (0.006)96.50.8793.7 (0.033)95.64.2

accuracy = (found concentration)/(nominal concentration) 100%; precision = (standard deviation)/(average of found concentration) 100%. conc. is concentration; acc. is accuracy; SD is standard deviation.