Research Article

Optimization of the Coupling Coefficient of the Inductive Link for Wireless Power Transfer to Biomedical Implants

Table 5

Performance comparisons with previous works of inductive link WPT system for cm-sized IMDs.

Reference[18][43][56][57]This work

TxRout (mm)14127.51526
Rin (mm)94.714
n152117
RxR2 (mm)0.552.9105
n274115
h (mm)1210162020
Frequency (MHz)2013.5630020025
RL (kΩ)0.250.5100.471
PTE, η (%)1.427.72.847.644.1
MPDL (mW)2.20.829.8145.4