Microwave Breast Imaging System Prototype with Integrated Numerical Characterization
Figure 10
Reconstructions of a single sphere (, ) = (40, 1) located at () = ( cm) of Example 1. (a) and (c) and (b) and (d) Relative permittivity and conductivity, respectively. (a) and (b) is the Born approximation. (c) and (d) is BIM iteration 4. Here, iterations help retrieve the relative permittivity in (c), but the Born approximation yielded better conductivity in (b).