Research Article

Rapid Determination of Mercury Ions in Environmental Water Based on an N-Rich Covalent Organic Framework Potential Sensor

Table 3

Recovery of the proposed GPE/COFTFPB-TZT electrode for the determination of Mercury ions in real water samples compared with the spectrophotometric method utilizing dithizone.

SampleSpikedSpectrophotometric method with dithizoneGPE/DTT electrodeRecovery (%)
(μmol/L)SD(μmol/L)SD

River water0.5000.502±0.0160.407±0.015101.8
River water0.6000.610±0.0130.491±0.01998.6
DI water0.7000.694±0.0140.604±0.014100.7
DI water0.8000.794±0.0130.797±0.02299.6
Pollution water1.1001.110±0.0151.106±0.014100.5
Pollution water1.6001.598±0.0111.596±0.01699.9