Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery
Table 2
EMG amplitude profiles of recurrent laryngeal nerve and vagus nerve stimulated by the attachable ring stimulator and conventional stimulator.
Patient no.
Side
V1 (μV)
R1 (μV)
R2 (μV)
V2 (μV)
Ring
Conventional
Ring
Conventional
Ring
Conventional
Ring
Conventional
1
Rt
2177
2077
100
2572
2406
166
2361
2460
99
2266
2262
4
Lt
1128
1177
49
1843
1733
110
2178
2530
352
1563
1191
372
2
Lt
1533
1003
530
1103
1773
670
1587
1635
48
996
1073
77
3
Lt
1735
1581
154
2599
2474
125
2658
2487
171
2182
2176
6
4
Lt
1021
1026
5
1375
1392
17
1216
1241
25
864
949
85
5
Rt
641
677
36
730
809
79
770
699
71
743
654
89
6
Rt
1031
977
54
2278
2188
90
1992
1799
193
1214
1111
103
7
Rt
3004
3043
39
2849
2809
40
2636
2583
53
2771
2539
232
Lt
1561
1505
56
1469
1518
49
1741
1584
157
1636
1774
138
8
Lt
762
808
46
541
566
25
834
850
16
561
571
10
9
Rt
621
640
19
716
801
85
1460
1356
104
1219
1157
62
10
Rt
660
653
7
788
786
2
862
902
40
690
689
1
Lt
1026
1058
32
956
981
25
650
603
47
668
634
34
11
Rt
2100
2159
59
2827
2735
92
3639
3633
6
2781
2595
186
12
Lt
607
628
21
1040
1038
2
2403
2040
363
1487
1255
232
13
Lt
849
697
152
1050
727
323
945
818
127
705
546
159
14
Rt
594
525
69
738
578
160
967
953
14
791
757
34
Lt
1146
1195
49
1666
1746
80
1391
1356
35
1177
1084
93
15
Rt
760
792
32
730
693
37
879
784
95
720
768
48
Lt
541
579
38
761
747
14
794
824
30
537
557
20
Mean
1175
1140
77
1432
1425
110
1598
1557
102
1279
1217
99
value†
0.867
0.979
0.876
0.782
∗Difference between the amplitudes evoked by the attachable ring stimulator and conventional stimulator. † value for the mean amplitudes between the two groups.