Research Article

Clinical Accuracy of Instrument-Read SARS-CoV-2 Antigen Rapid Diagnostic Tests (Ag-IRRDTs)

Table 2

Sensitivities extracted from 13 different Ag-IRRDT performance evaluation studies related to 15 cases for reference qRT-PCR values of Ct <30 (waived between 29 and 33) which yields an increase in pooled sensitivity value of the group from 0.73 to 0.85.

First author [ref]ManufacturerDescriptionCt <Sensitivity

V. Moeren [22]BD veritor<7 days300,983
V. Moeren [22]BD veritor>7 days300,854
Kweon [25]Boditech iChroma300,891
Baccani2 [49]Boditech AFIAS300,643
Kweon2 [25]Boditech AFIAS300,740
Krüger [31]LumiraDX300,902
Cento [34]LumiraDX290,908
Harris [38]Quidel sofiaSymptomatic, Ag+300,968
Harris [38]Quidel sofiaSymptomatic, Ag−300,630
Harris [38]Quidel sofiaAsymptomatic, Ag−300,077
Jaaskelainen [44]Quidel sofia300,943
FINDb [46]SD biosensor F330,809
FINDc [46]SD biosensor F330,750
Baccani [49]SD biosensor F300,714
Kahn [52]SD biosensor F300,642
Pooled Ct < 300,854