Research Article

Theoretical and Experimental Understanding of the Anomalous Odd-to-Even Isotope Ratios of Tin in a 1 + 1 Single-Colour Resonance Ionization Mass Spectrometry: Revisited

Table 3

A table of theoretically computed percentage anomalies for an inaccuracy of 0.5 and 1.0 for the various laser linewidths .

Laser linewidth (LLW) (GHz) % Anomaly ( ) for an inaccuracy of 0.5 LLW on either side of the peak frequency % Anomaly ( ) for an inaccuracy of 1.0 LLW on either side of the peak frequency% Anomaly ( ) @ resonance (0 GHz)
Inaccuracy in the laser peak frequency of 0.5 llw to the +ve side of the peak Inaccuracy in the laser peak frequency of 0.5 llw to the −ve side of the peakInaccuracy in the laser peak frequency of 1.0 llw to the +ve side of the peakInaccuracy in the laser peak frequency of 1.0 llw to the −ve side of the peak

1.2−5.63−4.92−5.35−3.94−5.5
1510.778.3329.3827.16−3.3
304.095.8212.8113.51−0.8
452.674.007.958.97−0.4
602.023.085.756.74−0.3