Research Article

One-Point and Multiline Calibration Fiber-Optic Laser-Ablation Spark-Induced Breakdown Spectroscopy for Quantitative Analysis of Elements in Aluminum Alloys

Table 3

The results of quantitative analysis of elements in four aluminum alloy samples by OP-MLC FO-LA-SIBS.

SampleElementMeasured concentration (wt.%)Certified concentration (wt.%)Relative errors (%)

1Mg0.92717.32
Cr0.1030.1531.26
Cu0.1540.1918.83
Mn0.0130.0132.25

2Mg0.7880.812.68
Cr0.0910.19.23
Cu0.1370.158.99
Mn0.0430.05116.36

4Mg0.4190.397.40
Cr0.0120.0120.24
Cu0.0430.05319.00
Mn0.1380.157.97

5Mg0.2270.218.12
Cr0.00710.004751.62
Cu0.0230.01640.85
Mn0.1690.2119.30