Research Article
An Optical Approach for the Determination of Droplet Volumes in Nanodispensing
Table 7
Gravimetric measurement of different quantities of drops
under constant conditions.
| Number of drops | 5000 | 7500 | 10 000 | 12 500 | 15 000 | Number of measurements | 15 | 15 | 15 | 15 | 15 | Mean value [mg or L] | 0.72 mg | 1.12 mg | 1.48 mg | 1.85 mg | 2.24 mg | Std. deviation | 0.0372 | 0.0373 | 0.0312 | 0.0398 | 0.0319 | % Std. deviation | 5.2% | 3.3% | 2.1% | 2.2% | 1.4% | Visual volume | 200 pl | 200 pl | 200 pl | 200 pl | 200 pl | Visual/gravimetric ratio | 1.40 | 1.33 | 1.35 | 1.35 | 1.34 |
|
|