Research Article

The Use of Amberlite Adsorbents for Green Chromatography Determination of Volatile Organic Compounds in Air

Table 3

VOCs recovery values expressed in % of signals found for deployed membranes as compared with spiked ones obtained for different adsorbents ( 𝑐 = 1 5 0 n g s a m p l e r 1 ).

CompoundRecovery %
VeramXAD-2XAD-4XAD-16

Chloroform1282624156
1,2-Dichloroethane63241645
Benzene43242635
Cyclohexane3110711180
Trichloroethylene95948852
Bromodichloromethane84151110
Toluene89556892
Tetrachloroethylene70474641
Dibromochloromethane70475056
Ethylbenzene73688399
m,p-Xylene9888131203
Bromoform105778784
o-Xylene60629764