Research Article

The Use of Amberlite Adsorbents for Green Chromatography Determination of Volatile Organic Compounds in Air

Table 4

Analytical features of HS-GC-MS determination of studied VOCs for each adsorbent evaluated.

CompoundVeramXAD-2XAD-4XAD-16
RSDLODRSDLODb RSDLODb RSDLODb
(%)ng · sampler−1(%)ng · sampler−1(×10−3)(%)ng · sampler−1(×10−3)(%)ng · sampler−1(×10−3)

Chloroform55013311.969302.049372.03
1,2-Dichloroethane95023300.9112730.8538130.87
Benzene710913.771043.661043.72
Cyclohexane550101300.8231300.7012391.08
Trichloroethylene650532.17441.9912231. 95
Bromodichloromethane6507101.324951.03141401.11
Toluene6105116.4847907.6122585.49
Tetrachloroethylene5104173.227412.9618292.85
Dibromochloromethane6504152.507281.7119191.64
Ethylbenzene910333.761263.221352.88
m,p-Xylene13106143.2811122.7917102.44
Bromoform1550120.991580.651970.75
o-Xylene1310282.8514212.47141142.13

Note: RSD %: relative standard deviation for 3 independent determinations ( 𝑐 = 1500 ng · sampler−1); LOD: Limit of detection expressed in ng · sampler−1; b: slope of calibration curve expressed in relative units.