Research Article
A Modified o-Phthalaldehyde Fluorometric Analytical Method for Ultratrace Ammonium in Natural Waters Using EDTA-NaOH as Buffer
Table 2
The FI of different concentration standard ammonium solution in existence and absence of EDTA.
| | 
 |  | Concentration of ammonium (, mol/L) | FI in existence of EDTA | FI in absence of EDTA |  | 
 |  | 0 | 62.871 | 74.994 |  | 0.125 | 253.749 | 235.408 |  | 0.250 | 444.613 | 414.782 |  | 0.375 | 589.166 | 551.247 |  | 0.500 | 760.241 | 690.585 |  | 
 |  | The relationship between FI and | FI = | FI = |  | 
 |  | Corresponding parameters |  |  |  |  | 0.9973 (  0.0001) | 0.9970 ( 0.0001) |  | The standard deviation of the intercept | 12.65 | 11.90 |  | The standard deviation of the slope | 41.32 | 38.87 |  | 
 | 
 |