Research Article

Determination of Selected Metals in Fruit Wines by Spectroscopic Techniques

Table 4

Basic validation parameters obtained for each analyte by using developed method (, number of standards in three replicates; , coefficient of determination).

AnalyteEquationLODLOQLinearity rangeCV

K60.9980.47 mg/L1.41 mg/L1.41–60 mg/L2.6
Ca50.9940.415 mg/L1.245 mg/L1.245–50 mg/L2.4
Mg50.9990.021 mg/L0.063 mg/L0.063–1.200 mg/L1.5
Pb50.9920.0031 µg/L0.009 µg/L0.0093–2.5000 µg/L2.0
Zn70.9980.027 µg/L0.081 µg/L0.081–1.500 µg/L1.9
Cd50.9970.0087 µg/L0.026 µg/L0.026–2 µg/L3.1
Fe50.9910.009 mg/L0.027 mg/L0.027–5 mg/L1.7
Hg50.9890.012 µg/L0.036 µg/L0.036–0.8 µg/L10.0
Sn50.9909.9 µg/L32.5 µg/L32.5–100 µg/Ln.d