A Highly Sensitive Multicommuted Flow Analysis Procedure for Photometric Determination of Molybdenum in Plant Materials without a Solvent Extraction Step
Table 4
Results of spiked recovery tests.
Sample
Analyte concentration (μg L−1)
Spiked concentration (μg L−1)
Found concentration (μg L−1)
Recovery%
1
74.3 ± 1.0
80
157.3 ± 0.6
103.7
2
94.3 ± 2.5
80
185.2 ± 3.6
113.5
3
73.1 ± 1.9
80
146.8 ± 2.3
92.2
4
73.9 ± 3.1
80
140.1 ± 2.6
82.7
5
73.9 ± 3.1
75
135.8 ± 1.6
82.5
6
73.1 ± 1.9
75
134.7 ± 3.4
82.1
7
74.3 ± 1.0
75
136.0 ± 1.3
82.2
Results are average of three consecutive measurements.