Research Article
Simple and Selective HPLC-UV/Vis Bioanalytical Method to Determine Aluminum Phthalocyanine Chloride in Skin Permeation Studies
Table 3
Intra- and interassay precision of the AlPcCl HPLC method.
| Theoretical concentration (μg/mL) | | | Experimental concentration (μg/mL) | CV (%) | Overall CV (%) |
| 1.0 | Analyst 1 | Day 1 | 1.00 | 1.49 | 7.80 | Equipment 1 | Day 2 | 1.03 | 2.71 | Analyst 2 | Day 1 | 0.88 | 3.74 | Equipment 2 | Day 2 | 0.83 | 3.07 | 2.5 | Analyst 1 | Day 1 | 2.49 | 1.12 | 1.72 | Equipment 1 | Day 2 | 2.54 | 4.33 | Analyst 2 | Day 1 | 2.49 | 0.88 | Equipment 2 | Day 2 | 2.49 | 1.94 | 5.0 | Analyst 1 | Day 1 | 5.05 | 1.39 | 1.66 | Equipment 1 | Day 2 | 5.03 | 4.26 | Analyst 2 | Day 1 | 4.53 | 2.58 | Equipment 2 | Day 2 | 4.48 | 1.82 |
|
|
CV: coefficient of variation.
|