Improvement in the Analytical Capabilities of LA-ICP-MS for High Spatial Resolution U-Pb Dating of Zircon Using Mixed-Gas Plasma
Table 1
Summary of the operating parameters used for LA-ICP-MS.
| GeoLas 2005 laser ablation system |
| Wavelength | ArF 193 nm | Pulse length | 15 ns | Energy density | 6–8 J·cm−2 | Spot size | 10, 16, 24, 32 μm | Repetition rate | 3 Hz | Ablation cell gas | Helium (0.5 l·min−1) | Makeup gas | Argon (0.631 l·min−1 in dry plasma, 0.72 l·min−1 in 2% ethanol plasma) | Ablation mode | Single spot |
| Agilent 7500a ICP-MS | RF forward power | 1380 W | Plasma gas flow rate | 14.0 l·min−1 | Auxiliary gas flow rate | 1.0 l·min−1 | Argon gas flow rate | 0.2 l·min−1 | Aqueous solution uptake rate | 0.33 ml·min−1 at 0.1 RPS | Skimmer cone | 0.4 mm Ni | Sampling cone | 1 mm Ni | Ion optical settings | Typical | Dwell time per isotope | 6 ms | Detector mode | Dual | Shield torch system | ON | Plasma conditions | Dry plasma or 2% ethanol plasma |
|
|