Research Article
Development of an Efficient Analytical Method for the Extraction and Analysis of Biocide Contents from the Textile Test Specimens on LC-DAD
Table 2
Results of robustness study on MIT.
| Condition | Assay, % | Retention time, minutes | Number of theoretical plates | Tailing | Resolution |
| 0.4% citric acid: methanol (50 : 50) | 100.18 | 2.506 | 18340 | 0.96 | 10.36 | 0.4% citric acid: methanol (55 : 45) | 99.42 | 2.485 | 18076 | 1.18 | 6.05 | 0.4% citric acid: methanol (45 : 55) | 100.60 | 3.365 | 18897 | 1.09 | 12.28 | Flow rate, 1.1 mL min−1 | 99.78 | 2.455 | 18225 | 1.06 | 9.85 | Flow rate, 0.9 mL min−1 | 100.33 | 2.535 | 17432 | 1.20 | 10.70 | Column oven temperature 33 | 100.40 | 2.475 | 16234 | 1.34 | 9.65 | Column oven temperature 27 | 100.38 | 2.511 | 18954 | 1.08 | 10.26 |
|
|