An Electrochemical Sensor Based on Gold Nanodendrite/Surfactant Modified Electrode for Bisphenol A Detection

Table 2

Determined BPA concentrations spiked in the real sample using AuNDs/CTAB/GCE and by Fluorescence methods.

SampleBPA (µm)
AddedFoundRecovery (%)
By AuNDs/CTAB/GCE sensorBy fluorescenceBy AuNDs/CTAB/GCE sensorBy fluorescence

Solution from LaVie plastics bottle0Not foundNot found
5.004.98 ± 0.064.88 ± 0.0799.5397.23
10.0010.28 ± 0.1410.23 ± 0.09102.83102.26
15.0014.49 ± 0.2914.89 ± 0.3396.6099.20

The recovery in each case is also shown.