Research Article
High-Accuracy Determination of Potassium and Selenium in Human Serum by Two-Step Isotope Dilution ICPMS
| | Parameter | K | Se |
| | ICP RF power (W) | 1100 | 1100 | | Gas flows (L/min) | 0.97 | 0.97 | | Lens voltage (V) | 6.75 | 6.75 | | Analog stage voltage (V) | −1700 | −1700 | | Pulse stage voltage (V) | 900 | 900 | | Dwell time per AMU | 2ms | 2ms | | Scan mode | Peak hopping | Peak hopping | | Sweeps/reading | 25 | 25 | | Readings/replicate | 20 | 25 | | Replicates | 15 | 25 | | Cell gas A : CH4 (mL/min) | 1.5 | 0.8 | | RP q: Ar (mL/min) | 0.8 | 0.45 |
|
|