Operational conditions for halogen determination by IC and by ICP-MS.
| | Parameter | Condition |
| | IC | | Stationary phase | Poly(vinyl alcohol) with quaternary ammonium groups | | Mobile phase (mmol L−1) | Na2CO3 (3.2)/NaHCO3 (1.0) | | Suppression solution (mol L−1) | H2SO4 (0.2) | | Flow rate (mL min−1) | 0.7 | | Sample loop (µL) | 20 | | Column | Metrosep A Supp 5 (150 mm x 4 mm i.d.) | | Guard column | Metrosep A Supp 4/5 Guard (5 mm × 4 mm i.d.) | | Detection | Conductivity |
| | ICP-MS | | Radiofrequency potency (W) | 1300 | | Plasma argon flux (L min−1) | 18 | | Auxiliary argon flux (L min−1) | 1.20 | | Nebulizer argon flux (L min−1) | 0.95 | | Sample uptake rate (mL min−1) | 0.7 | | Spray chamber | Cyclonic | | Nebulizer | Concentric | | Sampler and skimmer cones | Pt | | Ion lens | Auto lens “On” | | Isotopes (m/z) | 79Br, 127I |
|
|