Development and Validation of a Complexometric and Potentiometric Titration Method for the Quantitative Determination of Zinc Pyrithione in Shampoo
Table 2
Precision and accuracy (n = 6) of the method for potentiometric and complexometric titration.
| Parameter | Complexometric titration | Potentiometric titration |
| Intermediate precision RSD% | | 0.84 | 0.40 | Reproducibility RSD% | | 0.75 | 0.31 | 0.20% | 0.02 | 0.25 | System repeatability RSD% | 0.80% | 0.03 | 0.18 | 1.40% | 0.15 | 0.10 | Method repeatability RSD% | S1 | 0.47 | 0.30 | S2 | 0.70 | 0.23 | S3 | 0.69 | 0.36 | Accuracy % recovery | | 99.89 | 99.96 | Accuracy RSD % | | 1.06 | 0.98 |
|
|