Development and Validation of a Complexometric and Potentiometric Titration Method for the Quantitative Determination of Zinc Pyrithione in Shampoo

Table 2

Precision and accuracy (n = 6) of the method for potentiometric and complexometric titration.

ParameterComplexometric titrationPotentiometric titration

Intermediate precision RSD%0.840.40
Reproducibility RSD%0.750.31
0.20%0.020.25
System repeatability RSD%0.80%0.030.18
1.40%0.150.10
Method repeatability RSD%S10.470.30
S20.700.23
S30.690.36
Accuracy % recovery99.8999.96
Accuracy RSD %1.060.98