Research Article

A Novel Carbon Paste Electrode Modified with N1-Hydroxy-N1,N2-Diphenylbenzamidine for the Electrochemical Determination of Cadmium(II) in Environmental Samples

Table 4

Results of determination of Cd(II) in the real samples using the developed method (n = 5) and GFAAS method.

SampleAdded (nM)SWASV method detected (nM)SWASV method recovery (%)GFAAS method detected (nM)GFAAS method recovery (%)

Wastewater07.94 (±0.13)8.09 (±0.21)
3039.5 (±0.6)105.239.3 (±0.8)104.0

Bottled water0NDND
3030.8 (±0.7)102.731.3 (±0.4)104.3

Well water0NDND
3029.8 (±0.5)99.3329.6 (±0.7)98.66

Spring water0NDND
3028.8 (±0.2)96.0029.5 (±0.4)98.33
(ng·g−1)(ng·g−1)(ng·g−1)

Tobacco016.1 (±0.2)15.9 (±0.1)
3047.2 (±1.1)103.746.6 (±1.2)102.3

ND, not detected.