Research Article

Quantification of Myoinositol in Serum by Electrochemical Detection with an Unmodified Screen-Printed Carbon Electrode

Table 2

The precisions and recoveries of MI detected by the electrochemical method in human serum (n = 5).

Added concentration (μmol/L)Serum concentration (μmol/L)Measured concentration (mean ± SD, μmol/L)Precision (RSD, %)Recovery (%)

Intraday
 10.04.514.20 ± 0.453.2106
 1004.596.78 ± 5.225.496.7
 2004.5197.58 ± 12.266.296.6

Interday
 10.04.514.84 ± 1.087.3103
 1004.596.62 ± 8.639.090.3
 2004.5198.20 ± 14.047.195.2