Research Article
Quantification of Myoinositol in Serum by Electrochemical Detection with an Unmodified Screen-Printed Carbon Electrode
Table 2
The precisions and recoveries of MI detected by the electrochemical method in human serum (n = 5).
| Added concentration (μmol/L) | Serum concentration (μmol/L) | Measured concentration (mean ± SD, μmol/L) | Precision (RSD, %) | Recovery (%) |
| Intraday | 10.0 | 4.5 | 14.20 ± 0.45 | 3.2 | 106 | 100 | 4.5 | 96.78 ± 5.22 | 5.4 | 96.7 | 200 | 4.5 | 197.58 ± 12.26 | 6.2 | 96.6 |
| Interday | 10.0 | 4.5 | 14.84 ± 1.08 | 7.3 | 103 | 100 | 4.5 | 96.62 ± 8.63 | 9.0 | 90.3 | 200 | 4.5 | 198.20 ± 14.04 | 7.1 | 95.2 |
|
|