Research Article

Development and Validation of a Liquid Chromatography-Mass Spectrometry Assay for Determination of Cromolyn Sodium in Skin Permeation Studies

Table 1

Intraday and interday accuracy and precision for CS in 1X PBS.

Nominal concentration (μg/mL)
0.10.250.50.7512.5

Intraday
Day 1
Mean (N = 6)
0.0950.2430.4810.7241.0742.483
SD0.0160.0240.0610.0720.1170.215
Accuracy (% error)4.6322.9043.7323.4977.3780.680
Precision (% RSD)17.19.74312.7019.91410.9378.657
Day 2
Mean (N = 6)
0.0940.2210.4980.7611.0422.483
SD0.0160.0280.0640.0820.0730.141
Accuracy (% error)5.94611.4850.3421.4584.210.667
Precision (% RSD)17.53112.48212.90710.816.975.668
Day 3
Mean (N = 6)
0.1120.2820.4730.7320.9892.511
SD0.0220.020.0490.1050.1430.158
Accuracy (% error)12.10912.9645.3472.4661.0770.459
Precision (% RSD)19.4426.97910.29714.36814.4156.306

Interday
Average concentration (μg/mL)0.1010.2490.4840.7391.0352.493
SD0.0190.0340.0560.0840.1130.164
Accuracy (% error)15.98911.51110.189.2018.344.965
Precision (% RSD)19.12313.81711.55111.34510.9636.590