Research Article

Chemometric Methods for Simultaneous Determination of Candesartan Cilexetil and Hydrochlorothiazide in Binary Combinations

Table 10

Precision of the H-point standard addition method in the determination of HCT and CAN.

λRegression equationR2Added (µg/mL)Found (µg/mL)
HCTCANHCTCAN

283Y = 0.01103X + 0.58160.9944151515.1214.98
239Y = 0.03188X + 0.89390.9989

283Y = 0.01096X + 0.56880.9945151514.7514.86
239Y = 0.03194X + 0.88050.9985

283Y = 0.01141X + 0.56380.9917151514.4314.59
239Y = 0.03260X + 0.87300.9974

283Y = 0.01108X + 0.57190.9941151514.7115.06
239Y = 0.03187X + 0.88500.9989

283Y = 0.01103X + 0.57930.9944151515.1714.64
239Y = 0.03186X + 0.88430.9989

Mean14.8414.83
SD0.30840.2061
RSD(n=5)2.081.39
%R98.998.84

1 calculated using the regression equation of Y = 0.0247x + 0.0297 and the HCT calibration curve at 283 nm.