Research Article

Determination of the Levels of Selected Essential Metals in Sycamore (Ficus sycomorus L) Fruit and Seed Using Flame Atomic Absorption Spectrophotometry

Table 1

Instrumental operating conditions for determination of metals using F AAS.

ParametersKMgCaFeZnCuMn

λ (nm)404.6285.2422.7248.3213.9324.7279.5
S.W (nm)0.70.70.70.21.00.70.7

Flame typeAir-acetylene

L.C (mA)4.5123.50.21.53
Sample energy (eV)3.064.352.934.995.803.824.44

L.C = lamp current, nm = nanometre, mA = milliampere, eV = electron volt, S.W = slit width, λ = wavelength.