Research Article
Stress Detection Using Low Cost Heart Rate Sensors
Table 2
Time-domain results after the synchronization process.
| | Subject number | Mean RR (ms) | STD RR (ms) | | Schiller | CardioSport | Error | Schiller | CardioSport | Error |
| | #1 | 738.27 | 755.47 | 2.28% | 123.34 | 125.09 | 1.40% | | #2 | 704.04 | 720.42 | 2.27% | 91.35 | 93.47 | 2.27% | | #3 | 907.63 | 928.88 | 2.29% | 90.40 | 92.83 | 2.62% | | #4 | 854.53 | 874.50 | 2.28% | 144.74 | 148.00 | 2.20% | | #5 | 937.01 | 958.97 | 2.29% | 107.18 | 109.41 | 2.04% | | Average | 850.80 | 870.69 | 2.28% | 108.42 | 110.93 | 2.11% |
|
|