Research Article
Fully Automated Lumen Segmentation Method for Intracoronary Optical Coherence Tomography
Table 3
Statistical comparison of parameters between our methodology and commercially available system 1.
| | Our method (first measure) mean (CI) | System 1 (second measure) mean (CI) | Our method (first measure) median (IQR) | System 1 (second measure) median (IQR) | Difference | Relative difference | ICC (95% CI) | ICC value |
| Lumen area | 5.99 (5.83–6.14) | 6.04 (5.90–6.19) | 6.45 (4.72–7.67) | 6.50 (5.01–7.56) | −0.06 (−0.08–−0.03) | 1.67% (1.25%–2.10%) | 0.99 (0.98–0.99) | <0.0001 | Mean lumen diameter | 2.72 (2.68–2.76) | 2.72 (2.68–2.76) | 2.88 (2.41–3.13) | 2.87 (2.52–3.10) | −0.00 (−0.01–0.00) | 0.34% (0.06%–0.63%) | 0.98 (0.97–0.98) | <0.0001 | Minimal lumen diameter | 2.52 (2.48–2.56) | 2.52 (2.48–2.56) | 2.64 (2.18–2.96) | 2.52 (2.48–2.56) | 0.00 (−0.01–0.01) | 0.16% (−0.16%–0.49%) | 0.98 (0.97–0.98) | <0.0001 | Maximal lumen diameter | 2.91 (2.87–2.96) | 2.93 (2.89–2.97) | 3.07 (2.54–3.30) | 3.08 (2.66–3.28) | −0.02 (−0.03–−0.00) | 0.82% (0.43%–1.21%) | 0.94 (0.93–0.95) | <0.0001 |
|
|