Research Article
Fully Automated Lumen Segmentation Method for Intracoronary Optical Coherence Tomography
Table 4
Statistical comparison of parameters between our methodology and another commercially available system 2.
| | Our method (first measure) mean (CI) | System 2 (second measure) mean (CI) | Our method (first measure) median (IQR) | System 2 (second measure) median (IQR) | Difference | Relative difference | ICC (95% CI) | ICC value |
| Lumen area | 5.99 (5.83–6.14) | 5.76 (5.61–5.92) | 6.45 (4.72–7.67) | 6.13 (3.98–7.48) | 0.22 (0.18–0.27) | −4.23% (−4.88% to −3.57%) | 0.95 (0.93–0.97) | <0.0001 | Mean lumen diameter | 2.72 (2.68–2.76) | 2.66 (2.61–2.70) | 2.88 (2.41–3.13) | 2.79 (2.25–3.09) | 0.06 (0.05–0.07) | −2.42% (−2.85% to −1.99%) | 0.94 (0.92–0.96) | <0.0001 | Minimal lumen diameter | 2.52 (2.48–2.56) | 2.42 (2.37–2.46) | 2.64 (2.18–2.96) | 2.56 (2.01–2.91) | 0.10 (0.09–0.12) | −4.96% (−5.53% to −4.39%) | 0.95 (0.85–0.97) | <0.0001 | Maximal lumen diameter | 2.91 (2.87–2.96) | 2.87 (2.83–2.91) | 3.07 (2.54–3.30) | 3.00 (2.50–3.26) | 0.04 (0.02–0.06) | −1.11% (−1.58% to −0.63%) | 0.89 (0.88–0.91) | <0.0001 |
|
|