Research Article
[Retracted] Medical Equipment Comprehensive Management System Based on Cloud Computing and Internet of Things
Table 3
Monitor breakdown reason classification statistics table.
| Cause of issue | Number of failures |
| Accessories failure | Blood oxygen probe | 162 | ECG lead wire | 130 | Blood pressure cuff | 195 |
| Equipment failure | Blood oxygen module | 28 | Blood pressure module | 70 | ECG module | 28 | Power board | 11 | Fan | 43 | Display screen | 8 | Keyboard | 12 | Encoder | 46 | Main control board | 9 |
|
|