Interrater and Intrarater Reliability of Electrical Impedance Myography: A Comparison between Large and Small Handheld Electrode Arrays
Table 4
Reliability indices for between-days intrarater reliability of both sensors.
| Between-days intrarater | Large sensor | Small sensor | ICC | SEM | SRD | 95% LOA | ICC | SEM | SRD | 95% LOA |
| 50 kHz | Phase | 0.98 (0.95–0.99) | 0.13 | 0.99 | −5.66–3.81 | 0.84 (0.59–0.94) | 1.87 | 3.79 | −13.59–7.78 | Reactance | 0.90 (0.73–0.96) | 0.22 | 1.30 | −4.73–3.35 | 0.63 (0.02–0.85) | 3.30 | 5.04 | −15.28–8.62 | Resistance | 0.95 (0.85–0.98) | 0.54 | 2.03 | −11.73–6.85 | 0.95 (0.88–0.98) | 2.38 | 4.27 | −29.09–15.52 |
| 100 kHz | Phase | 0.98 (0.96–0.99) | 0.13 | 1.00 | −5.85–3.90 | 0.91 (0.76–0.96) | 0.99 | 2.76 | −10.10–6.30 | Reactance | 0.90 (0.74–0.96) | 0.22 | 1.31 | −4.76–3.36 | 0.84 (0.60–0.94) | 1.47 | 3.36 | −11.24–6.60 | Resistance | 0.96 (0.88–0.98) | 0.43 | 1.81 | −10.49–6.22 | 0.96 (0.89–0.98) | 1.98 | 3.90 | −25.98–13.97 |
|
|