Planning and Design of a Full-Outer-Air-Intake Natural Air-Conditioning System for Medical Negative Pressure Isolation Wards

Table 2

Design conditions of the full-outer-air-intake natural air-conditioning system for negative pressure isolation wards.

ItemNameData

1Ward bulk (m3)33.97
2Indoor temp. (°C)23∼25
3RH (%)50∼60
4CO2 concentration (ppm)600
5Negative pressure (Pa)8∼12
6Air exchange per hour12