Research Article
Multicenter Assessment of Radiation Exposure during Pediatric Cardiac Catheterizations Using a Novel Imaging System
Table 5
Study cohort median radiation exposure parameters and interquartile range (IQR) stratified by intervention type.
| Intervention | N | Air Kerma (mGy) from both planes | DAP (μGy m2) from both planes | DAP/kg (μGy m2/kg) from both planes | Fluoroscopy time (min) from both planes | Fraction Air Kerma from fluoroscopy | Median | IQR | Median | IQR | Median | IQR | Median | IQR | Median | IQR |
| ASD closure | 58 | 24 | 12–83 | 229 | 93–990 | 9.0 | 5–23 | 12 | 8–19 | 0.8 | 0.61–0.96 | PDA closure | 109 | 27 | 17–45 | 135 | 75–253 | 15 | 9–24 | 13 | 9–19 | 0.5 | 0.35–0.59 | Balloon dilation pulmonary valve | 45 | 29 | 11–47 | 116 | 47–326 | 21 | 11–31 | 16 | 10–21 | 0.4 | 0.2–0.52 | Balloon dilation aortic valve | 25 | 65 | 33–152 | 394 | 136–1601 | 34 | 17–53 | 15 | 12–28 | 0.4 | 0.15–0.46 | Coarctation intervention | 46 | 101 | 34–386 | 598 | 176–2649 | 45 | 22–90 | 19 | 11–27 | 0.4 | 0.2–0.44 | Transcatheter pulmonary valve | 17 | 889 | 539–1425 | 9869 | 6850–16616 | 197 | 153–249 | 51 | 30–62 | 0.5 | 0.26–0.68 |
|
|