Research Article

Long-Term Safety and Efficacy of Percutaneous Left Atrial Appendage Closure with the LAmbre Device

Table 2

Periprocedural data.

n = 56

LAA diameters
LAA length (mm)29.2 ± 6.5
LAA orifice diameter (mm)27.6 ± 5.2
LAA landing zone diameter (mm)22.6 ± 4.4
Number of LAA lobes
Single lobe42 (75)
Two lobes12 (21.4)
Multiple lobes2 (3.6)
Successful implantation56 (100)
LAA leak
Residual flow <1 mm1 (1.8)
Residual flow 1–3 mm5 (8.9)
Residual flow >3 mm0
Procedure time (min)60.1 ± 13.0
Contrast media (ml)44.5 ± 13.7
Complications
Death0
Stroke0
Pericardial effusion0
Major bleeding0
Major vascular complication0
Thrombosis with device0
Device dislocation0

LAA = left atrial appendage.