Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site

Table 2

Sample site comparison.

Sampling siteSuccessful measurementsMean ACT (s)SD (s)

Catheter9624453
Sheath9726280
IVL8622154

Matched comparisonsACT mean difference (s)SD value
Catheter vs. sheath (N = 94)−18.396.00.067
Catheter vs. IVL (N = 82)25.775.50.003
Sheath vs. IVL (N = 83)39.0102.80.001

ACT values obtained had a significantly lower ACT values as compared to both arterial catheter and arterial sheath. There were no significant differences between ACT values from arterial catheter and arterial sheath blood samples. Statistically significant difference in the two-tailed paired T-test (α = 0.05). s = seconds; SD = standard deviation; IVL = intravenous line.