Management of Failed Bioprosthetic Aortic Valves: Mitigating Complications and Optimizing Outcomes

Table 2

Surgical prosthesis amenable to valve fracture.

MakeStented/StentlessLeafletsFracture threshold (atm)Valve sizesID (mm)Profile height (mm)

CE magnaEdwards lifesciencesStentedInternal22–24
1918.014.0
2120.015.0
2322.016.0
2524.017.0
2726.018.0
2928.019.0

CE magna easeEdwards lifesciencesStentedInternal18
1918.013.0
2120.014.0
2322.015.0
2524.016.0
2726.017.0
2928.018.0

Perimount 2800/2900Edwards lifesciencesStentedInternal20
1918.014.0
2120.015.0
2322.016.0
2524.017.0
2726.018.0
2928.019.0

MitroflowSorin groupStentedExternal12
1915.411.0
2117.313.0
2319.014.0
2521.015.0
2722.916.0

MosaicMedtronicStentedInternal10
1917.513.5
2118.515.0
2320.516.0
2522.517.5
2724.018.5
2926.020.0

EpicAbbottStentedInternal8
1918.714.0
2120.815.0
2322.616.0
2524.517.0
2726.319.0

The Mosaic valve has been manufactured with two different materials and behaves differently during BVF depending on the material used to manufacture the frame. If the frame is made of Derlin, fracture occurs ∼10–12 atm. If comprised of the high-performance thermoplastic polyetheretherketone (PEEK) (a small amount in the Mosaic valve) it cannot be fractured but can me remodeled; continue to increase the inflation device pressure beyond 12 atm and at about 18 atm, the valve frame will begin to stretch. Inflate to ∼22 atm to achieve maximal expansion. Allen KB, Chhatriwalla AK, Saxon JT, et al. Bioprosthetic valve fracture: Technical insights from a multicenter study. J Thorac Cardiovasc Surg. 2019; 158 (5):1317–1328 e1311.