Research Article

Electrochemiluminescence Drug Sensor for Acetaminophen using Boron Nitride Quantum Dots as Ru(bpy)32+ Coreactant for Acetaminophen

Table 2

Recoveries of AP spiked from tap water samples (n = 3).

SampleStandard value of AP (M)Found (M)Recovery (%)RSD (%, n = 5)

Tap water1 × 10−69.23 × 10−7–9.87 × 10−792.3%–98.7%4.03
1 × 10−59.58 × 10−6–1.08 × 10−595.8%–108.0%3.74
5 × 10−55.08 × 10−5–5.32 × 10−5101.6%–106.4%4.26