Research Article

Accuracy of Intraocular Lens Power Formulas Involving 148 Eyes with Long Axial Lengths: A Retrospective Chart-Review Study

Table 4

Median absolute errors (D) calculated by each formula for Groups A–D.

FormulaGroup A
()
Group B
()
Group C
()
Group D
()

Haigis1.0800.8051.1602.145
Hoffer Q1.4201.6351.7103.430
Holladay1.2901.5301.4102.695
SRK/T1.0400.9750.9902.555
-value0.0020.0020.010/

Compared with the results achieved with Group D (repeated-measures ANOVA).
Group A: AL = 26.01–28.00 mm, Group B: AL = 28.01–30.00 mm, Group C: AL = 30.01–33.00 mm, and Group D: AL = 33.01–36.00 mm.
D, diopters.