Research Article

Morphometric Optic Nerve Head Analysis in Glaucoma Patients: A Comparison between the Simultaneous Nonmydriatic Stereoscopic Fundus Camera (Kowa Nonmyd WX3D) and the Heidelberg Scanning Laser Ophthalmoscope (HRT III)

Table 1

Comparison and distribution of stereometric optic nerve head parameters assessed with HRT III and Kowa nonmyd WX3D for all investigated patients subdivided in two groups according to their CDR (cut-off value 0.5) and in total. Values are expressed as mean (± standard deviation, SD).

Optic nerve head parameterHRT III ()Kowa nonmyd WX3D ()

CDR < 0.5 group
CDR0.36 (±0.10)0.50 (±0.12)
Disc area (mm2)2.18 (±0.32)2.83 (±0.57)
Cup volume (mm3)0.18 (±0.10)0.12 (±0.08)
Rim volume (mm3)0.38 (±0.16)0.50 (±0.35)
Maximum cup depth (mm)0.70 (±0.19)0.38 (±0.15)

CDR ≥ 0.5 group
CDR0.63 (±0.09)0.48 (±0.12)
Disc area (mm2)2.03 (±0.45)2.79 (±0.59)
Cup volume (mm3)0.30 (±0.16)0.11 (±0.08)
Rim volume (mm3)0.18 (±0.06)0.51 (±0.32)
Maximum cup depth (mm)0.72 (±0.25)0.38 (±0.17)

Total
CDR0.46 (±0.17)0.49 (±0.12)
Disc area (mm2)2.12 (±0.38)2.82 (±0.57)
Cup volume (mm3)0.22 (±0.14)0.11 (±0.08)
Rim volume (mm3)0.30 (±0.16)0.50 (±0.33)
Maximum cup depth (mm)0.70 (±0.21)0.38 (±0.15)