Comparison of Corneal Power and Astigmatism between Simulated Keratometry, True Net Power, and Total Corneal Refractive Power before and after SMILE Surgery

Table 2

Comparisons of the postoperative corneal cylindrical errors among simulated keratometry (simKa), true net power (TNPa), and total corneal refractive power (TCRPa) from 1 mm to 8 mm. All values were adjusted for the corneal plane using a vertex distance of 12 mm.

Vector meansDifference
simKaTNPaTCRPaTCRPa versus simKaTCRPa versus TNPasimKa versus TNPa
Magnitude|Angle|Magnitude|Angle|Magnitude|Angle|

1 mm(0.25 ± 0.62) × 75.2(0.21 ± 0.70) × 36.9(0.20 ± 0.69) × 39.20.27 ± 0.1910.5 ± 13.20.02 ± 0.091.3 ± 6.10.29 ± 0.189.9 ± 11.6
2 mm(0.49 ± 0.56) × 83.8(0.24 ± 0.61) × 69.7(0.24 ± 0.61) × 69.20.30 ± 0.1512.0 ± 14.90.01 ± 0.081.3 ± 8.20.30 ± 0.1511.2 ± 12.7
3 mm(0.71 ± 0.47) × 89.6(0.46 ± 0.48) × 88.6(0.46 ± 0.49) × 88.30.25 ± 0.1211.4 ± 15.70.01 ± 0.050.6 ± 0.90.25 ± 0.1211.2 ± 15.5
4 mm(0.92 ± 0.47) × 92.3(0.74 ± 0.49) × 93.2(0.75 ± 0.50) × 93.40.18 ± 0.126.1 ± 10.40.01 ± 0.040.6 ± 0.90.19 ± 0.125.8 ± 9.8
5 mm(1.07 ± 0.52) × 94.0(0.93 ± 0.51) × 94.6(0.98 ± 0.54) × 94.90.10 ± 0.133.5 ± 5.90.05 ± 0.050.6 ± 1.00.14 ± 0.123.2 ± 5.2
6 mm(1.16 ± 0.68) × 95.8(1.07 ± 0.67) × 95.7(1.17 ± 0.74) × 96.00.01 ± 0.143.3 ± 4.80.10 ± 0.080.8 ± 1.30.09 ± 0.112.7 ± 3.9
7 mm(1.15 ± 0.84) × 97.5(1.14 ± 0.85) × 96.5(1.31 ± 1.00) × 96.70.17 ± 0.254.5 ± 7.20.17 ± 0.171.1 ± 1.80.04 ± 0.153.7 ± 6.5
8 mm(1.09 ± 0.93) × 98.2(1.19 ± 0.98) × 96.4(1.44 ± 1.25) × 97.60.35 ± 0.475.8 ± 11.30.26 ± 0.393.7 ± 10.90.12 ± 0.152.9 ± 4.1

Magnitude represented the magnitude of the vector difference.
|Angle| was defined as the mean absolute values of the arithmetic difference between the axis of two corneal astigmatisms. It was defined always to be an acute angle.