Research Article

Disposable Carbon Dots Modified Screen Printed Carbon Electrode Electrochemical Sensor Strip for Selective Detection of Ferric Ions

Table 2

Percentage recovery of spiked Fe3+ ions for various water samples using C-Dots modified SPCE sensor strips and AAS.

Real sampleSample Fe(III) (ppm)Spiked Fe(III) (ppm)Measured Fe(III) (ppm)Recovery (%)RSD (%)Measured Fe(III) (ppm)Recovery (%)RSD (%)
AAS C-Dots modified SPCE sensor strip

Tap water<0.3 [25]109.010089.100.909.701894.025.16
Reverse osmosis waterND109.621596.210.109.744997.455.11
Mineral waterND108.601586.010.309.233692.346.85

ND = not detected.