Research Article
Accurate and Contactless Vital Sign Detection in Short Time Window with 24 GHz Doppler Radar
Table 3
Displacement MAE under different arc lengths.
| Amplitude | Central angle | MAE (percentage of error in displacement) | The proposed method | Ellipse fitting algorithm |
| 3 mm | 1.92π | 0.3049 mm (5.08%) | 0.3891 mm (6.49%) | 2.5 mm | 1.6π | 0.2044 mm (4.09%) | 0.2938 mm (5.88%) | 2 mm | 1.28π | 0.1944 mm (4.86%) | 0.1955 mm (4.89%) | 1.5 mm | 0.96π | 0.1502 mm (5.01%) | 0.3293 mm (10.98%) | 1 mm | 0.64π | 0.0827 mm (4.14%) | 0.1902 mm (9.51%) | 0.5 mm | 0.32π | 0.1008 mm (11.08%) | 2.5702 mm (257.02%) |
|
|